MILWAUKEE - ELECTROLYTIC OXYGEN SOLUTION 230ml for MW600 instrument
OUTDOOR" CALIBRATION PROCEDURE.
1. Remove the membrane
Remove the membrane from the probe end and dry the diode.
2. Reset the display
Connect the probe to the base unit and switch the unit on. The display should read 0.0. If it does not, go to the "0" position of the screw on the front of the unit with the screwdriver provided and through the access hole have the unit's display read 0.0.
3. Fill the membrane
Go to the membrane and put the "O" ring inside the membrane on the shoulder at the base of the wires. Fill the membrane just before the O-ring with the MA9071 electrolyte solution. Make sure there are no bubbles inside the membrane. (Air bubbles are the main cause of calibration failure)
4. Attach the membrane
Screw the membrane onto the probe.
5. Allow to stabilise
Switch on the instrument and hang the probe on the end of a table facing downwards and let the unit's display stabilise (this may take 5 to 15 minutes)
6. Adjust
Push down and hold down the 'Cal' button on the front of the instrument and use the screwdriver to go to the front Cal position '100%' and have the display show 100.
-